Basic Info.
Model NO.
HYTW-275C
Material
Stainless Steel
Brand
Huanyi
High Room Temperature Range
R.T(High Room Temperature) ~ +200ºC ≤60min
Temperature Rise Time
R.T~ -50ºC
Low Temperature Room Temperature Range
R.T(High Room Temperature) ~ +200ºC ≤60min
Temperature Drop Time
±0.5ºC
Temperature Shock Range
-40ºC~+150ºC
Temperature Fluctuation
≤±0.5ºC (No Load, Constant State)
Temperature Uniformity
≤2ºC (No Load, Constant State)
Temperature Departure
≤±2ºC (No Load, Constant State)
Transfer Time
≤10s
Transport Package
Wooden Packing
Specification
500*500*400mm
Trademark
HUANYI
Origin
China
HS Code
8514199000
Production Capacity
100/Month
Product Description
The Type Thermal Shock Test Chamber (two compartment temperature impact test chamber) can simulate the instantaneous changing environment between high and low temperatures,
used to determine whether equipment undergoes physical damage or performance degradation when subjected to rapid changes in ambient atmospheric temperature.
It is an essential testing equipment in the fields of aviation, automotive, home appliances, scientific research, and other fields.
It assesses and determines the parameters and performance of testing instruments, mechanical, electrical, electronic products, and components after high and low temperature impact tests.
Two chamber method high and low temperature impact test chamber, equipped with two studios, namely high temperature zone and low temperature zone.
The test piece is moved to the corresponding studio through a hanging basket for high and low temperature impact testin
The test piece is moved to the corresponding studio through a hanging basket for high and low temperature impact testin
The Two-box Type Thermal Shock Test Chamber meets the test standards
GB/T2423.1-2001 Test A: Low temperature test method
GB/T2423.2-2001 Test B: High temperature test method
GJB150.3A-2009 High temperature test
GJB150.4A-2009 Low temperature test
GJB150.5A-2009 Temperature shock test
Thermal shock test chamber specifications
Fault alarm and cause
Power-off protection function
Calendar timing function (automatic start and stop)
Self-diagnostic function